Measurement
• Measure pH, mV, or ISE with temperature (HI6000-2)
• Application-specific profiles allow quick and direct measurement without the need to
update the sensor and system settings
• Active log during measurement
• Measurement stability indicator (using the Stability Criteria setting)
• Reading modes:
• Direct and direct/autohold
• Known Addition, Known Subtraction, Analyte Addiiton, Analyte Subtration (HI6000-2)
• Temperature compensation can be Automatic (using temperature probe) or set manually
•Audible and/or alarm messages for measurements outside of predefined limits
•Galvanic isolation for pH/ORP measurement
Calibration
•pH calibration using
•Up to five Hanna Instruments pH buffers (pH 1.68, 3.00, 4.01, 6.86, 7.01, 9.18, 10.01 and
12.45)
•Up to five custom buffers
•mV calibration using a single point to calibrate offset.
•ISE calibration using up to five nominal standard values (e.g. for ppm: 0.010, 0.100, 1.00,
10.0, 100, 1000, 10000 ppm) and/or up to five custom solutions (user supplied)
Specifications
| SKU |
HI6000-1 |
| Name |
HI6000-1 pH/ORP Module designed to be used with the HI6000 Multiparameter Modular System. - HI6000-1 |
| pH Range |
-2.0 to 20.0 pH; -2.00 to 20.00 pH; -2.000 to 20.000 pH |
| pH Resolution |
0.1 pH; 0.01 pH; 0.001 pH |
| pH Accuracy |
±0.1 pH; ±0.01 pH; ±0.002 pH (±1 last significant digit) |
| Temperature
compensation |
Automatic or manual |
| Calibration
points |
up to 5 |
| Type |
Automatic; Semiautomatic; Manual |
| Standard
buffers |
Hanna and NIST pH 1.68, 3.00, 4.01, 6.86, 7.01, 9.18,
10.01, 12.45 |
| Custom buffers |
Up to 5 |
| Custom group |
Up to 5 |
| 1st calibration
point |
Offset or point (user setting) |
| Isopotential
point |
−2.000 to 20.000 pH |
| mV Isopotential
point |
−2.000 to 20.000 pH |
| mV Range |
−2000.0 mV to 2000.0 mV |
| mV Resolution |
1 mV; 0.1 mV |
| mV Accuracy |
±0.2 mV ±1 last significant digit |
| mV Calibration |
Single point offset, ±2000.0 mV |
| ISE Range |
1.0×10-5 to 300.0 ppt (g/L or
mg/mL)
5.0×10-3 to 1.0×105 ppm
(mg/L or µg/mL)
1.0 to 5.0×107 ppb (µg/L)
1.0×10-7 to 10.0 M (mol/L)
1.0×10-4 to 1.0×104 mmol/L
1.0×10-6 to 60.0 %w/v
5.0×10-7 to 5.0×107 user |
| ISE Resolution |
1, 2, 3 significant digits |
| ISE Accuracy |
±0.5% (monovalent ions)
±1% (divalent ions) |
| ISE Calibration points |
Up to 5 |
| ISE Calibration type |
All standards
Standard group |
| ISE Standards |
7 standard solutions available
for each concentration unit |
| ISE Custom standards |
Up to 5 |
| ISE Custom group |
Up to 5 |
| Temperature Range* |
-20.0 to 120.0 °C
-4.0 to 248.0 °F
253.2 to 393.2 K |
| Temperature Resolution |
0.1 °C
0.1 °F
0.1 K |
| Temperature Accuracy |
±0.2 °C
±0.4 °F
±0.2 K |
| Temperature Calibration |
Single point, adjustable |
| Reading mode |
Direct; Direct/Autohold |
| Reading mode Basic |
Measurement data
Measurement profile (if enabled)
Stability status |
| Reading mode Simple GLP |
Basic view information
Last calibration date, slope, offset (pH, Rel. mV - ISE
only) |
| Reading mode Full GLP |
Simple GLP information
Calibration point details (pH & ISE) |
| Reading mode Graph |
Basic view information
Measurement versus time graph |
| Reading mode Table |
Basic view information
Table with measurements updated every second |
Ordering Info
HI6000-02 (EU power plug) is supplied with HI764060
electrode holder; 24 VDC power adapter; USB-C to USB-A cable; quick reference guide and