| SKU |
HI83326-01 |
| Name |
Pool and Spa Photometer - HI83326 |
| pH Range |
Photometer: 6.5 to 8.5 pH
pH electrode: -2.00 to 16.00 pH |
| pH Resolution |
Photometer: 0.1 pH
pH electrode: 0.01 pH |
| pH Accuracy |
Photometer: ±0.1 pH
pH electrode: ±0.01 pH |
| pH Calibration |
Automatic one or two point calibration with one set of standard buffers available (4.01, 6.86, 7.01, 9.18, 10.01) |
| pH Temperature Compensation |
Automatic (-5.0 to 100.0 ℃; 23.0 to 212.0 °F); limits reduced based on the pH electrode used |
| pH CAL Check (electrode diagnostics) |
clean electrode and check buffer/check probe displayed during calibration |
| pH Method |
Photometer: phenol red |
| pH-mV Range |
±1000 mV |
| pH-mV Resolution |
0.1 mV |
| pH-mV Accuracy |
±0.2 mV |
| Absorbance Range |
0.000 to 4.000 Abs |
| Absorbance Resolution |
0.001 Abs |
| Absorbance Accuracy |
±0.003Abs @ 1.000 Abs |
| Alkalinity Range |
Freshwater: 0 to 500 mg/L (as CaCO3) |
| Alkalinity Resolution |
1 mg/L |
| Alkalinity Accuracy |
±5 mg/L ±5% of reading |
| Alkalinity Method |
Colorimetric method |
| Bromine Range |
0.00 to 8.00 mg/L (as Br2) |
| Bromine Resolution |
0.01 mg/L |
| Bromine Accuracy |
±0.08 mg/L ±3% of reading |
| Bromine Method |
Adaptation of the Standard Methods for the Examination of Water and Wastewater, 18th edition, DPD method. |
| Free Chlorine Range |
0.00 to 5.00 mg/L (as Cl2) |
| Free Chlorine Resolution |
0.01 mg/L |
| Free Chlorine Accuracy |
±0.03 mg/L ±3% of reading |
| Total Chlorine Range |
0.00 to 5.00 mg/L (as Cl2) |
| Total Chlorine Resolution |
0.01 mg/L |
| Total Chlorine Accuracy |
±0.03 mg/L ±3% of reading |
| Chlorine Method |
Adaptation of the EPA 330.5 DPD method |
| Copper Range |
High range: 0.00 to 5.00 mg/L (as Cu2+) |
| Copper Resolution |
0.01 mg/L |
| Copper Accuracy |
High Range ±0.02 mg/L ±4% of reading |
| Copper Method |
Adaptation of the EPA bicinchoninate method |
| Cyanuric Acid Range |
0 to 80 mg/L (as CYA) |
| Cyanuric Acid Resolution |
1 mg/L |
| Cyanuric Acid Accuracy |
±1 mg/L ±15% of reading |
| Cyanuric Acid Method |
Adaptation of the turbidimetric method |
| Hardness, Calcium Range |
0.00 to 2.70 mg/L (as CaCO3) |
| Hardness, Calcium Resolution |
0.01 mg/L |
| Hardness, Calcium Accuracy |
±0.11 mg/L ±5% of reading |
| Hardness, Calcium Method |
Adaptation of the Standard Methods for the Examination of Water and Wastewater, 18th edition, Calmagite method |
| Iron Range |
High Range: 0.00 to 5.00 mg/L (as Fe) |
| Iron Resolution |
0.01 mg/L |
| Iron Accuracy |
High Range: ±0.04 mg/L ±2% of reading |
| Iron Method |
High Range: Adaptation of Standard Methods for the Examination of Water and Wastewater, 3500-Fe B., Phenanthroline Method |
| Nitrate Range |
0.0 to 30.0 mg/L (as NO3-- N) |
| Nitrate Resolution |
0.1 mg/L |
| Nitrate Accuracy |
±0.5 mg/L ±10% of reading |
| Nitrate Method |
Adaptation of the cadmium reduction method |
| Ozone Range |
0.00 to 2.00 mg/L (as O3) |
| Ozone Resolution |
0.01 mg/L |
| Ozone Accuracy |
±0.02 mg/L ±3% of reading |
| Ozone Method |
Colorimetric DPD Method |
| Phosphate Range |
Freshwater
Low Range: 0.00 to 2.50 mg/L (as PO43-) |
| Phosphate Resolution |
Freshwater: 0.01 mg/L |
| Phosphate Accuracy |
Freshwater
Low Range: ±0.04 mg/L ±4% of reading |
| Phosphate Method |
Freshwater Low Range: Adaptation of the Ascorbic Acid method |
| Input Channels |
1 pH electrode input and 3 photometer wavelengths |
| pH Electrode |
digital pH electrode (not included) |
| Photometer/Colorimeter Light Source |
3 LEDs with 525 nm, 575 nm, and 610 nm narrow band interference filters |
| Photometer/Colorimeter Light Detector |
silicon photodetector |
| Bandpass Filter Bandwidth |
8 nm |
| Bandpass Filter Wavelength Accuracy |
±1 nm |
| Cuvette Type |
round, 24.6 mm |
| Number of Methods |
128 max. |
| GLP |
calibration data for connected pH electrode |
| Display |
128 x 64 pixel LCD with backlight |
| Logging Type |
log on demand with user name and sample ID optional input |
| Logging Memory |
1000 readings |
| Connectivity |
USB-A host for flash drive; micro-USB-B for power and computer connectivity |
| Power Supply |
5 VDC USB 2.0 power adapter with USB-A to micro-USB-B cable (included) |
| Battery Type/Life |
3.7 VDC Li-polymer rechargeable battery / >500 photometric measurements or 50 hours of continuous pH measurement |
| Environment |
0 to 50.0 ℃ (32 to 122.0 °F); 0 to 95% RH, non-condensing |
| Weight |
1.0 kg (2.2 lbs.) |
| Dimensions |
206 x 177 x 97 mm (8.1 x 7.0 x 3.8 in.) |